HFCAS OpenIR  > 中科院合肥智能机械研究所
Relative validity of an indirect calorimetry device for measuring resting energy expenditure and respiratory quotient
Wang, Xi1,2; Wang, Yuan1; Ding, Zenghui1,2; Cao, Guangbei1; Hu, Fusong1,2; Sun, Yining1; Ma, Zuchang1; Zhou, Duoqi3; Su, Benyue3
2018
发表期刊ASIA PACIFIC JOURNAL OF CLINICAL NUTRITION
ISSN0964-7058
摘要

Background and Objectives: Resting energy expenditure (REE) and respiratory quotient (RQ) are important for optimal nutritional care. The purpose of this study was to assess the accuracy and repeatability of an indirect calorimetry device (IIM-IC-100) in the measurement of REE and RQ in healthy Chinese adults. Methods and Study Design: A total of 38 healthy adults (19 male and 19 female) aged 18-52 years (25 +/- 6 years) were enrolled in this study. REE and RQ were measured by IIM-IC-100 and by VO2000, alternately and in duplicate. Results: There was a highly significant correlation between IIM-IC-100 REE and VO2000 REE (r=0.906, p<0.001), with mean IIM-IC-100 REE significantly higher than that of VO2000 (1475 +/- 269 vs 1394 +/- 313 kcal/d, p=0.002). Bland-Altman analysis revealed that the mean difference between IIM-IC-100 REE and VO2000 REE was 81.3 kcal/d, with limits of agreement of -185 to + 347 kcal/d. There was no significant difference in RQ between the two devices. No significant differences were observed between the repeated measurements for both devices. Intra-subject coefficients of variation (CVs) of REE were smaller for IIM-IC-100 (5.8%) than for VO2000 (10.5%), while CVs of RQ were similar for IIM-IC-100 (7.2%) and VO2000 (6.9%). Conclusions: These preliminary data indicated that the IIM-IC-100 showed promise as an accurate and precise tool in the assessment of REE and RQ in healthy Chinese adults.

关键词indirect calorimetry resting energy expenditure respiratory quotient oxygen consumption validation
DOI10.6133/apjcn.032017.02
关键词[WOS]BODY-MASS INDEX ; METABOLIC-RATE ; CLINICAL VALIDATION ; VENTILATED CANOPY ; CHINESE ADULTS ; GAS-EXCHANGE ; FACE MASK ; OBESITY ; WOMEN ; RELIABILITY
收录类别SCI
语种英语
资助项目National Natural Science Foundation of China[61603371] ; Major projects of science and technology in Anhui Province[15czz02019] ; Science and Technology Service Network Initiative[KFJ-SW-STS-161] ; Science and Technology Service Network Initiative[KFJ-SW-STS-161] ; Major projects of science and technology in Anhui Province[15czz02019] ; National Natural Science Foundation of China[61603371]
WOS研究方向Nutrition & Dietetics
WOS类目Nutrition & Dietetics
WOS记录号WOS:000419140900006
出版者H E C PRESS, HEALTHY EATING CLUB PTY LTD
引用统计
被引频次:6[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.hfcas.ac.cn:8080/handle/334002/35006
专题中科院合肥智能机械研究所
通讯作者Ma, Zuchang
作者单位1.Chinese Acad Sci, Inst Intelligent Machines, Res Ctr Informat Technol Sports & Hlth, Hefei 230031, Anhui, Peoples R China
2.Univ Sci & Technol China, Dept Automat, Hefei, Anhui, Peoples R China
3.Anqing Normal Univ, Sch Phys Educ, Anqing, Anhui, Peoples R China
推荐引用方式
GB/T 7714
Wang, Xi,Wang, Yuan,Ding, Zenghui,et al. Relative validity of an indirect calorimetry device for measuring resting energy expenditure and respiratory quotient[J]. ASIA PACIFIC JOURNAL OF CLINICAL NUTRITION,2018,27(1):72-77.
APA Wang, Xi.,Wang, Yuan.,Ding, Zenghui.,Cao, Guangbei.,Hu, Fusong.,...&Su, Benyue.(2018).Relative validity of an indirect calorimetry device for measuring resting energy expenditure and respiratory quotient.ASIA PACIFIC JOURNAL OF CLINICAL NUTRITION,27(1),72-77.
MLA Wang, Xi,et al."Relative validity of an indirect calorimetry device for measuring resting energy expenditure and respiratory quotient".ASIA PACIFIC JOURNAL OF CLINICAL NUTRITION 27.1(2018):72-77.
条目包含的文件 下载所有文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Relative validity of(577KB)期刊论文作者接受稿开放获取CC BY-NC-SA浏览 下载
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Wang, Xi]的文章
[Wang, Yuan]的文章
[Ding, Zenghui]的文章
百度学术
百度学术中相似的文章
[Wang, Xi]的文章
[Wang, Yuan]的文章
[Ding, Zenghui]的文章
必应学术
必应学术中相似的文章
[Wang, Xi]的文章
[Wang, Yuan]的文章
[Ding, Zenghui]的文章
相关权益政策
暂无数据
收藏/分享
文件名: Relative validity of an indirect calorimetry device for measuring resting energy expenditure and respiratory quotient.pdf
格式: Adobe PDF
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。