HFCAS OpenIR  > 中科院安徽光学精密机械研究所
High Sensitivity Detection of Carbon Dioxide Based on Portable Cylindrical Multi-Pass Cell
Chen Jia-jin1,2; Wang Gui-shi1; Liu Kun1; Tan Tu1; Cheng Gang1; Tian Xing1; Gao Xiao-ming1,2
2019
发表期刊SPECTROSCOPY AND SPECTRAL ANALYSIS
ISSN1000-0593
摘要

Carbon dioxide is one of the most important greenhouse gases playing an important role in climate change and human activities, so the detection of carbon dioxide concentration has an important significance. Highly sensitive detection of carbon dioxide gas was achieved by using near infrared tunable diode laser absorption spectroscopy combined with self-designed portable miniaturized cylindrical mirror multi-pass absorption cell. The cylindrical mirrors multi-pass absorption cell was optimized and designed by using a light transmission matrix programmed with Matlab software, compared with traditional Herriott multipass cell, which has the advantages of high mirror utilization and longer optical path length in the same volume. An effective optical path length of 14 m was achieved with 15 cm physical length. In present work, a DFB diode laser emission at 1. 57 mu m was used as a light source. Direct absorption spectroscopy method was used to detect the CO2, and the Allan variance was used to analyze the system performance. The results showed that the detection sensitivity of the system can be achieved 5. 3 mu L.L-1 with the average time of 5 s, and a detection sensitivity of 33. 1 mu L.L-1 can be achieved by averaging in 235 s. In addition, CO2 in the atmosphere was measured by the developed CO2 sensor, the measured results showed that the concentration of CO2 in the atmosphere is about 383. 4 mu L.L-1 The tunable diode laser absorption spectroscopy (IDLAS) system based on cylindrical mirrors multi-pass absorption cell, combined with the cylindrical mirrors multi pass cell can achieve long optical path in a small volume and tunable diode laser absorption spectroscopy technology has the advantages of high sensitivity, high resolution and fast response, greatly reduces volume and improves detection sensitivity of the system, which has a broad application in the field of gas detection.

关键词Tunable laser Direct absorption spectroscopy Optical multi-pass absorption cell Carbon dioxide Detection
DOI10.3964/j.issn.1000-0593(2019)01-0292-05
关键词[WOS]DIODE-LASER ; CO2 ; SPECTROSCOPY
收录类别SCI
语种英语
WOS研究方向Spectroscopy
WOS类目Spectroscopy
WOS记录号WOS:000456357000045
出版者OFFICE SPECTROSCOPY & SPECTRAL ANALYSIS
引用统计
被引频次:1[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.hfcas.ac.cn:8080/handle/334002/41457
专题中科院安徽光学精密机械研究所
通讯作者Wang Gui-shi; Gao Xiao-ming
作者单位1.Chinese Acad Sci, Anhui Inst Opt & Fine Mech, Hefei 230031, Anhui, Peoples R China
2.Univ Sci & Technol China, Hefei 230026, Anhui, Peoples R China
第一作者单位中科院安徽光学精密机械研究所
通讯作者单位中科院安徽光学精密机械研究所
推荐引用方式
GB/T 7714
Chen Jia-jin,Wang Gui-shi,Liu Kun,et al. High Sensitivity Detection of Carbon Dioxide Based on Portable Cylindrical Multi-Pass Cell[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS,2019,39(1):292-296.
APA Chen Jia-jin.,Wang Gui-shi.,Liu Kun.,Tan Tu.,Cheng Gang.,...&Gao Xiao-ming.(2019).High Sensitivity Detection of Carbon Dioxide Based on Portable Cylindrical Multi-Pass Cell.SPECTROSCOPY AND SPECTRAL ANALYSIS,39(1),292-296.
MLA Chen Jia-jin,et al."High Sensitivity Detection of Carbon Dioxide Based on Portable Cylindrical Multi-Pass Cell".SPECTROSCOPY AND SPECTRAL ANALYSIS 39.1(2019):292-296.
条目包含的文件 下载所有文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
基于便携式柱面镜多通池的二氧化碳高灵敏度(817KB)期刊论文作者接受稿开放获取CC BY-NC-SA浏览 下载
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Chen Jia-jin]的文章
[Wang Gui-shi]的文章
[Liu Kun]的文章
百度学术
百度学术中相似的文章
[Chen Jia-jin]的文章
[Wang Gui-shi]的文章
[Liu Kun]的文章
必应学术
必应学术中相似的文章
[Chen Jia-jin]的文章
[Wang Gui-shi]的文章
[Liu Kun]的文章
相关权益政策
暂无数据
收藏/分享
文件名: 基于便携式柱面镜多通池的二氧化碳高灵敏度探测研究_陈家金.pdf
格式: Adobe PDF
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。