摘要为测量红外模拟探测器量子效率,搭建了基于参量下转换效应的测量系统.利用光学参量振荡器调谐输出的2 576.5nm波长连续激光束作为泵浦源、ZnGeP2晶体作为非线性晶体,获取了3.42μm和10.4μm的相关光子对.实验中通过将采集到的两个通道的光电流/电压信号转换为等效光子计数,并利用两光路平衡测量的方法对每个模式的平均光子计数进行校正,实现了碲镉汞红外模拟探测器在10.4μm处的绝对功率响应度定标,得到了相对合成不确定度优于4.53%的定标结果.
其他摘要The measurement facility based on parametric down-conversion technique is established to measure the quantum efficiency of the infrared analogy detector. Uses the CW laser beam of 2 576.5 nm wavelength tuning out of Optical Parametric Oscillator (OPO) laser as the pump source and the ZnGeP2 crystal as the nonlinear crystal, obtained the correlated photon pairs of 3.42μm and 10.4μm. The charge amount of the collected photo-current/voltage signal is converted into the equivalent photon counting, and the dual channel data fluctuation is carried out for the average photon counting of each mode through the method of double light path balance measurement, realized the absolute power responsivity calibrated of an MCT infrared detector at 10.4μm, the results show a relative combination uncertainty of about better than 4.53 % for this calibration method.
关键词遥感 参量下转换 相关光子 模拟探测器 量子效率定标
收录类别CSCD
语种中文
CSCD记录号CSCD:6102443
引用统计
文献类型期刊论文
条目标识符http://ir.hfcas.ac.cn:8080/handle/334002/68748
专题中国科学院合肥物质科学研究院
推荐引用方式
GB/T 7714
. 碲镉汞模拟探测器量子效率定标方法[J]. 光子学报,2017,046.
APA (2017).碲镉汞模拟探测器量子效率定标方法.光子学报,046.
MLA "碲镉汞模拟探测器量子效率定标方法".光子学报 046(2017).
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
百度学术
百度学术中相似的文章
必应学术
必应学术中相似的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。