HFCAS OpenIR
FTIR光谱仪中傅里叶插值采样方法的研究
其他题名Fourier interpolative sampling algorithm in Fourier transform infrared spectrometer
2018
发表期刊红外与激光工程
ISSN1007-2276
摘要为解决传统"Brault"采样方法中参考信号过零点计算复杂、耗时较多的问题,提出了基于傅里叶插值技术寻找过零点的方法。通过与其他插值方法进行比较研究,结果表明,这种方法确保了过零点信息准确性的同时简化了数据处理的复杂度,得到的过零点信息线性拟合系数大于0.999。在2 100~2 200 cm^-1波段范围内,当参考激光信号误差较小时,傅里叶插值方法得到的仪器SNR是三次样条插值法所得到的1.03倍,而线性插值方法与傅里叶插值方法得到的结果一致;当参考激光信号误差较大时,傅里叶插值方法得到的仪器SNR是线性插值方法得到的仪器SNR的1.05倍。
其他摘要In order to solve the problem of complex and time-consumption of reference signal zero crossings in the traditional "Brault" sampling method, a method based on Fourier interpolation technique was proposed to find the zero-crossing. Compared with other interpolation methods, the results showed that this method could ensure the accuracy of zero-crossing information and simplify the complexity of data processing. The linear fitting coefficient of zero-crossing information obtained was greater than 0.999. In the range of 2 100-2 200 cm^-1, when the error of reference laser signal was small, the instrumental SNR obtained by the Fourier interpolation method was 1.03 times that obtained by the cubic spline interpolation method, and the result obtained by the linear interpolation method was consistent with Fourier interpolation method. When the error of reference laser signal was relatively large, the instrumental SNR obtained by the Fourier interpolation method was 1.05 times of the SNR obtained by the linear interpolation method.
关键词采样方法 傅里叶变换红外光谱仪 “Brault” 方法 傅里叶插值 过采样
收录类别CSCD
语种中文
CSCD记录号CSCD:6178106
引用统计
被引频次:1[CSCD]   [CSCD记录]
文献类型期刊论文
条目标识符http://ir.hfcas.ac.cn:8080/handle/334002/68948
专题中国科学院合肥物质科学研究院
推荐引用方式
GB/T 7714
. FTIR光谱仪中傅里叶插值采样方法的研究[J]. 红外与激光工程,2018,047.
APA (2018).FTIR光谱仪中傅里叶插值采样方法的研究.红外与激光工程,047.
MLA "FTIR光谱仪中傅里叶插值采样方法的研究".红外与激光工程 047(2018).
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
百度学术
百度学术中相似的文章
必应学术
必应学术中相似的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。